Detect reverse flow in MFC #20

Open
sjorsdewit opened this issue Apr 13, 2020 · 0 comments
Open

Detect reverse flow in MFC #20

sjorsdewit opened this issue Apr 13, 2020 · 0 comments

Comments

@sjorsdewit
Copy link
Member

If the oxygen/air supply hose is disconnected (or leaking), reverse flow may occur through the MFCs. It appears that the MFC feedback is an absolute quantity of flow: we may read +5SLPM, but in case of disconnected hose, this could be -5SLPM..!?